Description and specifications
5L/min micro dry pump
Product name: 5 L / min micro dry pump
Model: SE-5, SE-E1-5, SE-E2M-5
Power: 20 W
Applications: Portable mass spectrometers, gas sampling, biological culture applications, etc.
Product Description: Scroll Vacuum Pump, Oil-free Scroll Pump, Laboratory Vacuum Pumps, Oil-free vacuum pumps, Micro Dry Pumps
Technical specifications
| Parameter | Specification |
|---|---|
| Pumping speed | 5 L / min |
| Motor power | ≤ 20 W |
| Motor type | brushless DC motor |
| Supply voltage | 12 VDC |
| Motor speed | 3600 RPM |
| Water vapor treatment capacity | small amounts of liquid water can be removed |
| Intake pressure | 1 bar (gauge pressure) |
| Exhaust pressure | 0.6 bar (gauge pressure) |
| Noise (OEM) | 56 dB (A) |
| Noise (with enclosure) | 50 dB (A) |
| Weight (OEM) | 1.1 kg |
| Weight (with enclosure) | 1.3kg |
| Size (OEM) | 183 x 87 x 93mm |
| Dimensions (with enclosure) | 185 x 132 x 119 mm |
| Inlet / outlet connector (with enclosure) | G 1 / 4" |
| Inlet / outlet connector (bare pump) | G 1 / 8" |
| Cooling | Air cooling |
| Working temperature | −20 °C to 40 °C |
Model variants and ultimate pressure
| Parameter | Specification |
|---|---|
| SE-E0-5 | Ultimate pressure ≤ 1 mbar (100 Pa) |
| SE-E1-5 | Ultimate pressure ≤ 0.1 mbar (10 Pa) |
| SE-E2M-5 | Ultimate pressure ≤ 0.05 mbar (5 Pa) |
Design and operating features
- Has all the advantages of traditional scroll dry pumps;
- No top sealing structure, no PTFE grinding chips, no need to replace the seal regularly;
- Basically without fragile parts, free from daily maintenance, long service time;
- It can function normally at wider ambient temperatures (traditional scroll is only 12-40 ° C);
- High airtightness, a higher vacuum can be obtained, and the pumping capacity is stronger when approaching the top of the vacuum;
- High reliability, with strong handling capabilities for water vapour, alcohol, acetone, etc.
- Provide a clean vacuum, oil-free operation, and the surrounding environment is not polluted;
- They can operate in multiple spaces without installation location restrictions.
The base model designation differs between the overview and the variant table; confirm the exact model when ordering.